On August 18, 2026, the FDA's Center for Devices and Radiological Health released a 30-page discussion paper titled Considerations for the Regulation of Generative AI-Enabled Medical Devices. It is not guidance — the paper says so on page one — and it does not propose policy changes. It is the first time the agency has put in writing how it might regulate medical devices built on large language models. The docket is FDA-2026-N-7874 on Regulations.gov.

Acting Commissioner Kyle Diamantas framed the effort as part of the push to accelerate AI-powered medical products to market. CDRH Director Dr. Michelle Tarver called the process "a potential model for regulators around the world."

The FDA's AI-Enabled Medical Device List has grown to 1,524 entries as of March 2026, with 333 cleared in 2025 alone. Roughly 97% entered through the 510(k) pathway, which requires substantial equivalence to a previously cleared predicate. More than three-quarters are radiology tools — narrow, single-task systems trained to flag a lesion or optimize a scan. Generative AI does not fit this model. These devices accept open-ended inputs, produce variable outputs to similar queries, and are often built on third-party foundation models whose training data and update cycles are opaque to the sponsor. As Ashkon Rasooli, founder of EnGenius Solutions, told MD+DI in January: "There are currently no pathways to get a generative AI-enabled medical device on the market."

The Two-Axis Grid and the Doctor's Exam

The paper's core proposal is a two-axis risk framework. One axis maps the degree of device autonomy — from non-directive information (a cardiovascular risk score) through action-directing output (a recommendation to seek emergency care) to fully autonomous action. The other maps the severity of harm from an incorrect output. Risk rises from the lower-left to the upper-right. FDA's own examples are specific: hydrocortisone advice for poison ivy sits low; whether to go to the ER for chest pain sits high even though both are "just information." Autonomously starting a thrombolytic order set for stroke is higher than autonomously prescribing antibiotics for confirmed strep. The evidence expectation follows the position on the grid, not the presence of an LLM.

For premarket evaluation, the paper proposes what it calls a competency-based approach — modeled, FDA says, on how human clinicians are evaluated and credentialed. The structure has two stages: non-clinical device benchmarking, then clinical confirmation in settings of increasing realism. Clinicians are not tested on every scenario they might encounter, FDA reasons, and neither can a device with open-ended inputs.

The benchmarking framework defines ten elements across four groups: safety (critical recognition and escalation, scope maintenance, calibration and uncertainty), clinical proficiency (knowledge and task fidelity, information gathering, quantitative analysis, communication), generalizability (robustness and reproducibility, subgroup performance), and agentic conduct (planning within safety envelopes, human checkpoints before irreversible actions). The paper explicitly permits an LLM to serve as an independent adjudicator — "those independence expectations would still be applicable when the expert adjudicator is itself an LLM" — provided it meets the same standards as a human reviewer.

Clinical confirmation ranges across five rungs of increasing rigor: retrospective evaluation on real patient inputs, shadow deployment with outputs hidden from care, standardized patient interactions with trained actors, clinician adjudication of real cases, and prospective studies including RCTs. The paper states it "might not require a prospective clinical study in every case." The evaluation target is "the final user-facing device, as configured and intended to be deployed for real-world use, and not the foundation model standing alone."

The Postmarket Trade and the Model Behind the Device

Section VI opens with the sentence that could reshape the economics of this category: "CDRH is considering whether it is appropriate to accept greater premarket uncertainty regarding a GenAI-enabled device's benefit-risk profile through greater reliance on postmarket monitoring." The monitoring options include periodic re-benchmarking against premarket thresholds, sample-based clinician review of real-world interactions, and drift monitoring to detect performance degradation as models, prompts, or components change. FDA asks whether "machine-based supervisory agents" could run some of the monitoring — an AI watching the AI. The premarket competency assessment becomes the baseline: a modified device is re-benchmarked against the same capabilities, with a Predetermined Change Control Plan as one mechanism.

The foundation model problem gets its own section. Many GenAI devices are built on third-party models whose developers control refusal behavior, content policies, and safety-critical features while offering limited transparency. FDA floats a voluntary Foundation Model Device Master File — built on its existing Device Master File program — where model developers could submit architecture, training data provenance, healthcare failure modes, subgroup benchmark results, and guardrails, held confidentially and referenced by sponsors with the holder's permission. A Master File would not authorize the model for any use. The sponsor still proves the device. But it gives FDA a way to inspect the component that may matter most to device behavior. Question 25 asks: given "limited incentive to disclose," what would make a voluntary Master File useful?

Agentic AI — systems that "autonomously plan and execute multi-step tasks, use external tools, or take actions across a sequence of steps" — gets a definition and a benchmarking element of its own (A.1), covering human checkpoints before irreversible actions and injection through retrieved content. The paper notes that documentation and outreach agents may not be device functions at all.

The EU AI Act's high-risk obligations take effect this month for medical device AI, adding a parallel layer for companies in both markets. Between fiscal years 2024 and 2025, FDA's own internal AI use cases grew 148%. What the framework becomes depends on what arrives in the docket by October 19.